Abstract
For decades spinal surgeons have attempted to design simple, single stage anterior internal fixation systems for the thoracic and lumbar spine. Early devices presented both biomechanical and technical problems. The AO Anterior Thoracolumbar Locking Plate (ATLP) was designed to solve some of the problems encountered with early anterior instrumentation. The ATLP system is constructed in Commercially Pure titanium. It is a low profile device indicated for use for unstable burst fractures in the anterior column; metastatic tumor management; and degenerative diseases of the thoracolumbar spine between levels T10 and L5. Implantation of the device involves direct anterior decompression with sagittal reduction and corpectomy. This is followed by grafting reconstruction, and plate fixation. This device has been implanted in 25 patients with an average follow-up of 38 months. There were five (5) broken screws in three (3) patients, and no broken plates. Implant related postoperative complications included two misplaced screws. Preliminary results indicate that the ATLP system seems to be a safe, low profile, MRI/CT compatible device that provides definitive single stage fixation of the anterior spinal column.
Similar content being viewed by others
Log in or create a free account to read this content
Gain free access to this article, as well as selected content from this journal and more on nature.com
or
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Thalgott, J., Kabins, M., Timlin, M. et al. Four year experience with the AO Anterior Thoracolumbar Locking Plate. Spinal Cord 35, 286–291 (1997). https://doi.org/10.1038/sj.sc.3100399
Issue date:
DOI: https://doi.org/10.1038/sj.sc.3100399
Keywords
This article is cited by
-
Biomechanical evaluation of position and bicortical fixation of anterior lateral vertebral screws in a porcine model
Scientific Reports (2023)
-
Biomechanical analysis of a new expandable vertebral body replacement combined with a new polyaxial antero-lateral plate and/or pedicle screws and rods
European Spine Journal (2012)