Table 2 Mean and regression statistics for aPWV measures plotted against SphygmoCor aPWV

From: Validity and repeatability of the Vicorder apparatus: a comparison with the SphygmoCor device

    

Linear regression

Device and distance method

aPWV (m s−1)

Difference (m s−1)

Significance (P)

Slope

r

P

SphygmoCor (transit time=60±16 ms)a

 Notch-femoral – notch-carotid

8.6±2.9

0

1

1

0

<0.001

Vicorder (transit time=83±18 ms)

 Notch-femoral (manufacturer)

8.3±2.2

−0.3±1.5

0.14

0.64

0.85

<0.001

 Notch-femoral – notch-carotid

7.0±1.9

1.6±1.6

<0.001

0.57

0.86

<0.001

+ Cuff pulse distance

7.9±2.1

0.7±1.5

<0.001

0.63

0.86

<0.001

Vicorder, femoral segment removed (transit time=65±18 ms)

–Femoral segment

7.6±2.8

1.0±1.6

<0.001

0.83

0.84

<0.001

  1. Abbreviation: aPWV, aortic pulse wave velocity.
  2. aReference value.