Table 1 Measured aberrations of the probe-forming system in the SrTiO3 experiment.

From: Atomic electric fields revealed by a quantum mechanical approach to electron picodiffraction

Aberration type

Value

A2 (threefold astigmatism)

55±40 nm

B2 (coma)

30±20 nm

C3 (spherical aberration Cs)

−0.8±1.9 μm

A3 (fourfold astigmatism)

2.7±1.1 μm

S3 (star aberration)

600±650 nm