Abstract
We have developed two devices: a high-amplitude bubble continuous positive airway pressure (HAB-CPAP) and an inexpensive bubble intermittent mandatory ventilator (B-IMV) to test the hypotheses that simple, inexpensive devices can provide gas exchange similar to that of bubble CPAP (B-CPAP) and conventional mechanical ventilation (CMV). Twelve paralyzed juvenile rabbits were intubated, stabilized on CMV, and then switched to CPAP. On identical mean airway pressures (MAPs), animals were unable to maintain pulse oximeter oxygen saturation (SpO2) >80% on conventional B-CPAP, but all animals oxygenated well (97.3 ± 2.1%) on HAB-CPAP. In fact, arterial partial pressures of O2 (Pao2) were higher during HAB-CPAP than during CMV (p = 0.01). After repeated lung lavages, arterial partial pressures of CO2 (Paco2) were lower with B-IMV than with CMV (p < 0.0001), despite identical ventilator settings. In lavaged animals, when HAB-CPAP was compared with CMV at the same MAP and 100% O2, no differences were observed in Pao2, but Paco2 levels were higher with HAB-CPAP (70 ± 7 versus 50 ± 5 mm Hg; p < 0.05). Arterial blood pressures were not impaired by HAB-CPAP or B-IMV. The results confirm that simple inexpensive devices can provide respiratory support in the face of severe lung disease and could extend the use of respiratory support for preterm infants into severely resource-limited settings.
Similar content being viewed by others
Log in or create a free account to read this content
Gain free access to this article, as well as selected content from this journal and more on nature.com
or
Abbreviations
- ABG:
-
arterial blood gas analyses
- B-CPAP:
-
bubble CPAP
- B-IMV:
-
bubble intermittent mandatory ventilation or ventilators
- B-IMV0:
-
B-IMV with PEEP bubbler angle of 0
- B-IMV90:
-
B-IMV with PEEP bubbler angle of 90
- B-IMV135:
-
B-IMV with PEEP bubbler angle of 135
- CPAP:
-
continuous positive airway pressure
- CMV:
-
conventional mechanical ventilation
- FiO2:
-
fraction of inspired oxygen
- HAB-CPAP:
-
high-amplitude bubble CPAP
- HAB-CPAP0:
-
HAB-CPAP with bubbler angle of 0°, functionally equivalent to conventional B-CPAP
- HAB-CPAP90:
-
HAB-CPAP with bubbler angle of 90°, parallel with the surface of water
- HAB-CPAP135:
-
HAB-CPAP with bubbler angle of 135°
- IMV:
-
intermittent mandatory ventilation
- MAP:
-
mean airway pressures
- Paw:
-
airway pressures
- PEEP:
-
positive end-expiratory pressure
- SNK:
-
Student-Newman-Keuls
- SpO2:
-
pulse oximeter oxygen saturations.
References
Singh GK, Yu SM 1995 Infant mortality in the United States: trends, differentials, and projections 1950 through 2010. Am J Public Health 85: 957–964
Heron M, Sutton PD, Xu J, Ventura SJ, Strobino DM, Guyer B 2010 Annual summary of vital statistics: 2007. Pediatrics 125: 4–15
Lawn JE, Cousens S, Zupan J 2005 4 million neonatal deaths: when? Where? Why?. Lancet 365: 891–900
Singhal N, Bhutta ZA 2008 Newborn resuscitation in resource-limited settings. Semin Fetal Neonatal Med 13: 432–439
DiBlasi RM, Zignego JC, Tang DM, Hildebrandt J, Smith CV, Hansen TN, Richardson CP 2010 Noninvasive respiratory support of juvenile rabbits by high-amplitude bubble continuous positive airway pressure. Pediatr Res 67: 624–629
Zar JH 1984 Biostatistical Analysis. Prentice-Hall, Englewood Cliffs, NJ
Reyburn B, Li M, Metcalfe DB, Kroll NJ, Alvord J, Wint A, Dahl MJ, Sun J, Dong L, Wang ZM, Callaway C, McKnight RA, Moyer-Mileur L, Yoder BA, Null DM, Lane RH, Albertine KH 2008 Nasal ventilation alters mesenchymal cell turnover and improves alveolarization in preterm lambs. Am J Respir Crit Care Med 178: 407–418
Versmold HT, Brunstler I, Schlosser C 1982 High Frequency Oscillation Durch Blubber-CPAP Beschleunigt CO2 Elimination. Thieme, Stuttgart, Germany, pp 159–162
Randell SH, Mercer RR, Young SL 1990 Neonatal hyperoxia alters the pulmonary alveolar and capillary structure of 40-day-old rats. Am J Pathol 136: 1259–1266
Randell SH, Mercer RR, Young SL 1989 Postnatal growth of pulmonary acini and alveoli in normal and oxygen-exposed rats studied by serial section reconstructions. Am J Anat 186: 55–68
Pillow JJ, Travadi JN 2005 Bubble CPAP: is the noise important? An in vitro study. Pediatr Res 57: 826–830
Author information
Authors and Affiliations
Additional information
Supported by the Seattle Children's Research Institute.
Patent applications concerning the devices described in this manuscript have been submitted to USPTO (US Provisional Patent Application Numbers 61/044,002 61/150,670) and to International Patent Application (PCT/US2009/039957) by Seattle Children's Research Institute (SCRI). The applications are in review. Authors R.M.D., J.C.Z., C.V.S., T.N.H., and C.P.R. are inventors on the patent applications and could benefit financially from the inventions. http://www.wipo.int/pctdb/en/wo.jsp?WO=2009126739.
Rights and permissions
About this article
Cite this article
Diblasi, R., Zignego, J., Smith, C. et al. Effective Gas Exchange in Paralyzed Juvenile Rabbits Using Simple, Inexpensive Respiratory Support Devices. Pediatr Res 68, 526–530 (2010). https://doi.org/10.1203/PDR.0b013e3181f985f0
Received:
Accepted:
Issue date:
DOI: https://doi.org/10.1203/PDR.0b013e3181f985f0
This article is cited by
-
Bubble bilevel ventilation facilitates gas exchange in anesthetized rabbits
Pediatric Research (2021)
-
A trial comparing continuous positive airway pressure (CPAP) devices in preterm infants
Journal of Perinatology (2020)
-
Evaluating the efficacy of Seattle-PAP for the respiratory support of premature neonates: study protocol for a randomized controlled trial
Trials (2019)
-
Continuous positive airway pressure in preterm neonates: An update of current evidence and implications for developing countries
Indian Pediatrics (2015)
-
Bubble vs Conventional Continuous Positive Airway Pressure for Prevention of Extubation Failure in Preterm Very Low Birth Weight Infants: A Pilot Study
The Indian Journal of Pediatrics (2012)