Table 2 Participant characteristics and reference blood pressure distribution.

From: Validation testing of five home blood pressure monitoring devices for the upper arm according to the ISO 81060-2:2018/AMD 1:2020 protocol

Blood Pressure Monitor Model

 

BP7450

BP5450

UA-8000WM

WGNBPA-240BT

BP3MW1-4YCVS

Male (N): Female (N)

30: 55

26: 60

26: 59

26: 59

26: 59

Age (years)

  Range (Low: High)

18: 86

21: 87

21: 83

21: 86

21: 86

  Mean (SD)

47.2 (20.3)

48.9 (20.3)

44.0 (17.8)

47.3 (19.5)

45.6 (18.8)

Arm Circumference (cm)

  Range (Low: High)

22: 42

18.5: 42

24: 42

23: 42

22: 42

  Mean (SD)

31.8 (5.4)

30.7 (6.7)

32.9 (5.3)

32.8 (5.3)

32.2 (5.6)

SBP (mmHg) Overall Range (Low: High)

83.5: 207.5

80: 202.5

84: 202.5

85: 201

85: 198

≥160 (N: %)

16: 6.3

15: 5.8

14: 5.5

13: 5.1

14: 5.5

≥140

59: 23.1

52: 20.2

52: 20.4

51: 20.0

51: 20.0

≤100

54: 21.2

55: 21.3

62: 24.3

53: 20.8

53: 20.8

DBP (mmHg) Overall Range (Low: High)

50: 114

46.5: 113

43.5: 115.5

45.5: 111.5

45: 114.5

≥ 100 (N: %)

16: 6.3

13: 5.0

16: 6.3

15: 5.9

16: 6.3

≥ 85

57: 22.3

54: 20.9

51: 20.0

51: 20.0

51: 20.0

≤ 60

40: 15.7

43: 16.7

41: 16.1

46: 18.0

45: 17.6