Table 3 Variables for noise floor calculation in the proposed system with QCM
Variables | 3-DoF | 5-DoF |
|---|---|---|
Noise spectral density \({{\rm{N}}}_{1}\)(@1 Hz) | \(21.3{\rm{\mu }}{\rm{V}}/\sqrt{{\rm{Hz}}}\) | \(22.2{\rm{\mu }}{\rm{V}}/\sqrt{{\rm{Hz}}}\) |
Noise spectral density \({{\rm{N}}}_{{\rm{n}}}\)(@1 Hz) | \(9.8{\rm{\mu }}{\rm{V}}/\sqrt{{\rm{Hz}}}\) | \(5.6{\rm{\mu }}{\rm{V}}/\sqrt{{\rm{Hz}}}\) |
Measured signal \({{\rm{X}}}_{1}\) | 0.802 V | 0.805 V |
Measured signal \({{\rm{X}}}_{{\rm{n}}}\) | 0.337 V | 0.087 V |
Scale factor \({{\rm{K}}}_{{\rm{scale}},{\rm{out}}1}\) | \({2}^{15}\) | \({2}^{21}\) |
Scale factor \({{\rm{K}}}_{{\rm{scale}},{\rm{out}}2}\) | \({2}^{6}\) | \({2}^{12}\) |
Normalized sensitivity | 505638 | 3238894 |