Table 5 Variables for noise floor calculation in the proposed system with DETF
Variables | 3-DoF | 5-DoF |
|---|---|---|
Noise spectral density \({{\rm{N}}}_{1}\)(@1 Hz) | \(176.4{\rm{\mu }}{\rm{V}}/\sqrt{{\rm{Hz}}}\) | \(178.2{\rm{\mu }}{\rm{V}}/\sqrt{{\rm{Hz}}}\) |
Noise spectral density \({{\rm{N}}}_{{\rm{n}}}\)(@1 Hz) | \(190.3{\rm{\mu }}{\rm{V}}/\sqrt{{\rm{Hz}}}\) | 76.9\({\rm{\mu }}{\rm{V}}/\sqrt{{\rm{Hz}}}\) |
Measured signal \({{\rm{X}}}_{1}\) | 0.556 V | 0.559 V |
Measured signal \({{\rm{X}}}_{{\rm{n}}}\) | 0.625 V | 0.219 V |
Scale factor \({{\rm{K}}}_{{\rm{scale}},{\rm{out}}1}\) | \({2}^{14}\) | \({2}^{20}\) |
Scale factor \({{\rm{K}}}_{{\rm{scale}},{\rm{out}}2}\) | \({2}^{6}\) | \({2}^{11}\) |
Normalized sensitivity | 537288 | 3308454 |