Table 2 Aortic pressure as obtained by catheter, brachial oscillometric device, GTF method, and mathematical modeling

From: Multiscale mathematical modeling vs. the generalized transfer function approach for aortic pressure estimation: a comparison with invasive data

    

Errors

  

Mean

Std. deviation

Mean

Std. deviation

R

Aortic invasive pressure

SBP [mm Hg]

141.3

20.2

   
 

DBP [mm Hg]

76.0

12.2

   
 

PP [mm Hg]

65.3

17.8

   

Brachial non-invasive pressure

SBP [mm Hg]

144.0

17.8

-2.7

12.0

0.808

 

DBP [mm Hg]

81.5

11.7

-5.5

8.6

0.742

 

PP [mm Hg]

62.5

13.8

2.8

12.5

0.715

GTF method

SBP [mm Hg]

133.1

17.3

8.2

10.0

0.869

 

DBP [mm Hg]

82.4

12.0

-6.4

8.7

0.743

 

PP [mm Hg]

50.7

13.2

14.7

10.6

0.807

Mathematical modeling

SBP [mm Hg]

137.0

21.6

4.3

16.7

0.679

 

DBP [mm Hg]

72.2

16.7

3.8

10.4

0.784

 

PP [mm Hg]

64.8

19.5

-0.5

18.7

0.500