Table 1 LODs, LOQs, and cutoffs using pGold chip and CIA.

From: Diagnosis and prognosis of myocardial infarction on a plasmonic chip

 

cTnI (ng mL−1)

CK-Mb (ng mL−1)

CIA-LODa

0.0152

1.3245

CIA-LOQb

0.0208

2.8393

CIA-cutoffs

0.04

4.00

pGold-LODa

0.0100

0.2715

pGold-LOQb

0.0157

1.1348

pGold-cutoffs

0.03

3.13

  1. aLODs were calculated by applying the mean blank value plus 3 times of s.d. to the fitting line of calibration curves.