Table 2 Conditions for turbulent flow chromatography (TFC) loading and eluting pumps for analysis of SARS-CoV-2 nucleoprotein target peptides.

From: Establishing a mass spectrometry-based system for rapid detection of SARS-CoV-2 in large clinical sample cohorts

Step

Start (min)

Loading pump

Eluting pump

Flow

Grad

%A

%B

%C

%D

Tee

Loop

Flow

Grad

%A

%B

1

0

1.2

Step

100

====

out

0.3

Step

99

1

2

0.5

1.2

Step

100

====

out

0.3

Step

99

1

3

1

0.2

Step

100

====

out

0.3

Step

99

1

4

1.1

0.1

Step

100

T

in

0.3

Step

99

1

5

2.7

1.2

Step

60

40

====

in

0.3

Step

80

20

6

3.9

1.2

Step

100

====

out

0.3

Ramp

75.5

24.5

7

4.9

1.2

Step

100

====

out

0.3

Ramp

75.0

25

8

5.9

1.2

Step

100

====

out

0.3

Ramp

72.5

27.5

9

6.9

1.2

Step

100

====

out

0.3

Ramp

70

30

10

8.4

0.5

Step

100

====

out

0.3

Ramp

50

50

11

8.6

0.5

Step

100

====

out

0.3

Step

20

80

12

9.6

1.2

Step

100

====

out

0.3

Step

99

1

  1. Steps 1 to 3: sample loading; step 4: transfer of peptides onto the analytical column; steps 5 to 10: switching wash between organic solvents (C and D) to clean large particles and reduce carryover in the TFC column, and peptide elution from the analytical column; step 11: TFC column equilibration and analytical column washing; step 12: system equilibration.
  2. Loading: A (1‒4) 0.5% formic acid in water; B (1‒4) acetonitrile; C (1‒4) isopropanol, acetonitrile, and acetone (40/40/20, v/v/v); D (1–4) 20% dimethyl sulfoxide (DMSO) and 2% trifluoroethanol in water.
  3. Eluting: A (1‒4) 1% DMSO, 0.1% formic acid in water; B (1‒4) 1% DMSO, 0.1% formic acid in acetonitrile.