Fig. 2: Reader study results. | Nature Communications

Fig. 2: Reader study results.

From: Artificial intelligence system reduces false-positive findings in the interpretation of breast ultrasound exams

Fig. 2

The performance of the AI system on the reader study population (n = 1024 breasts) using ROC curve (a) and precision-recall curve (b). The AI achieved 0.962 (95% CI: 0.943, 0.979) AUROC and 0.752 (95% CI: 0.675, 0.849) AUPRC. Each data point represents a single reader and the triangles correspond to the average reader performance. The inset shows a magnification of the gray shaded region.