Fig. 4: Detection of MBP-CoV-2NP. | Nature Communications

Fig. 4: Detection of MBP-CoV-2NP.

From: A Fabry-Pérot cavity coupled surface plasmon photodiode for electrical biomolecular sensing

Fig. 4

a Offset photocurrent at fixed angle as a function of time. The sensor is initially covered with a buffer solution. At time = 0, the buffer solution is removed and replaced with solutions containing different concentrations of antigen (MBP-CoV-2NP). At time = 2 min the antigen solution is removed and replaced with buffer solution. b Change of photocurrent after 1 min as a function of antigen concentration. Symbols and error bars are the mean and SD of four different sensors. The dotted red line is the noise floor of the sensor as defined by the standard deviation of the current during the initial buffer solution stage (time = −0.5 to 0 min). The dashed black line is a best-fit guide to the eye. Source data are provided as a Source Data file.