Table 3 Detection accuracy and repeatability (Conc.mean/ σconc./ Err.) of the crown ether decorated SiP Pb2+ sensor in environmental-related scenarios (tap, lake, sea water), using the respective samples, where the Pb2+ concentrations are verified by ICP-MS
From: Crown ether decorated silicon photonics for safeguarding against lead poisoning
Field samples | Inferred Conc.mean/ σconc. (n = 6)/ Err. |
---|---|
Tap water, undoped; 2 ppb of Pb2+, measured by ICP-MS | 2.40 ppb/ 0.3 ppb/20.0% |
Lake water, undoped; 5 ppb of Pb2+, measured by ICP-MS | 4.77 ppb/ 0.6 ppb/4.6% |
Sea water, undoped; 8 ppb of Pb2+; measured by ICP-MS | 8.16 ppb/ 0.4 ppb/2.0% |
Doped tap water, where Pb2+ concentration is synthetically increased by 15 ppb to 17 ppb, measured by ICP-MS | 16.7 ppb/ 0.5 ppb/1.8% |
Doped lake water, where Pb2+ concentration is synthetically increased by 15 ppb to 20 ppb, measured by ICP-MS | 19.5 ppb/ 0.7 ppb/2.5% |
Doped sea water, where Pb2+ concentration is synthetically increased by 15 ppb to 23 ppb, measured by ICP-MS | 22.8 ppb/ 0.7 ppb/0.9% |