Table 1 Comparative performance of four LIT modes, Luminex, and ELISA

From: Lab-in-a-Tip: a multiplex immunoassay platform based on a self-assembled barcoded protein array

 

Standard LIT

Speedy LIT

Microvolume LIT

Ultrasensitive LIT

ELISA

Luminex

Sensitivity

pg/ml

pg/ml

pg/ml

fg/ml

pg/ml

pg/ml

Dynamic range

4 logs

4 logs

4 logs

3 logs

3 logs

4 logs

Throughput

Multiplex

Multiplex

Multiplex

Multiplex

Single-plex

Multiplex

Incubation time

1 h

15 min

1 h

1.5 h

5 h

3.5 h

Reaction volume

50 μl

50 μl

10 μl

50 μl

100 μl

50–100 μl