Table 1 All estimated noise parameters for the data acquired with different primary ion beam currents
From: Orbitrap noise structure and method for noise unbiased multivariate analysis
Primary ion beam current (pA) | A | RN | σW | σF | \(\bar{\sigma }\) | A / \(\bar{\sigma }\) | Excluded m/q |
|---|---|---|---|---|---|---|---|
50 | 19.79(27) | 1.747(40) | 10.81(4) | 1.53(2) | 13.2 | 1.49 | 317 <m/q < 359 |
100 | 21.78(22) | 2.021(12) | 11.77(4) | 1.81(2) | 14.8 | 1.47 | 281<m/q < 792 |
200 | 25.72(36) | 3.143(59) | 13.07(7) | 2.27(3) | 17.2 | 1.49 | 251<m/q < 792 |