Extended Data Fig. 4: Detailed measurement setup. | Nature

Extended Data Fig. 4: Detailed measurement setup.

From: Bolometer operating at the threshold for circuit quantum electrodynamics

Extended Data Fig. 4

The blue rectangle corresponds to the SEM image displayed in Fig. 1a. The model numbers on the filters, couplers and amplifiers refer to Mini-Circuits product numbers. Nominal temperatures of the different cryostat plates (dashed horizontal lines) are indicated. The 180° bends and spirals refer to long segments in the coaxial microwave cables and Thermocoax cables, respectively, which function as thermal bottlenecks separating the different temperature stages. The top-most dashed line refers to the boundary of the cryostat. The nested millikelvin radiation shields are made of gold-plated copper.