Extended Data Fig. 7: Magnetic and seismic coupling.
From: Measurement of gravitational coupling between millimetre-sized masses

a, The time-resolved amplitude spectral density of the horizontal ground motion in Vienna, recorded by an STS2 broadband seismometer, shows a strong frequency dependence of the variability. A rise of the noise floor is observed throughout the frequency range, starting from midnight (dark blue) to noon (light blue), but is most prominent at 30–40 mHz. By contrast, the noise floor at the microseismic peak, at around 70 mHz, varies only slightly. The modulation frequency of the drive mass is marked by a vertical dashed line. Each spectrum corresponds to a single 2-h measurement. b, The amplitude spectral density of magnetic field variations at the site of the experiment were measured with a three-axis magnetometer (Stefan Mayer FLC3-70; z = vertical, y = along source–test mass axis). The signal at 10 mHz is explained by the period of the traffic light at a nearby crossing, regulating car traffic as well as trams. The inferred force acting on the pendulum (red line), shows no sign of this signal. Each spectrum corresponds to a single 8,000-s measurement.