Table 6 Statistics of the “check” standard quality-control metrics as presented in Fig. 6 for the three instrument types.
From: A 120,000-year long climate record from a NW-Greenland deep ice core at ultra-high resolution
Instrument | δ18O precision [‰] | δ18O accuracy [‰] | δD precision [‰] | δD accuracy [‰] |
---|---|---|---|---|
L2120-i | 0.065 ± 0.028 | 0.023 ± 0.059 | 0.524 ± 0.163 | 0.151 ± 0.232 |
L2130-i | 0.031 ± 0.017 | −0.001 ± 0.035 | 0.301 ± 0.098 | 0.034 ± 0.146 |
L2140-i | 0.034 ± 0.015 | −0.011 ± 0.033 | 0.390 ± 0.118 | 0.012 ± 0.172 |