Table 2 Average degradation rates of Mg-based helical stent under in vitro.
From: Expandable Mg-based Helical Stent Assessment using Static, Dynamic, and Porcine Ex Vivo Models
 | Static immersion | Dynamic artificial vascular bioreactor* | ||
---|---|---|---|---|
Expansion rate (%) | 0 | 0 | ~58.8 | ~98.5 |
Degradation rate (mm/y) | 0.99 ± 0.20 | 2.44 ± 0.45 | 1.62 ± 0.17 | 1.64 ± 0.39 |