Figure 3 | Scientific Reports

Figure 3

From: Measurement of local temperature increments induced by cultured HepG2 cells with micro-thermocouples in a thermally stabilized system

Figure 3

Calibration results of the Pd/Cr and Cr/Pt TFTCs. The calculated thermopower is the average value of 9 different TFTCs. (a) Calibration results of 9 different Pd/Cr TFTCs. These TFTCs have varied stripe width of 2 μm, 20 μm, 50 μm, 100 μm, 200 μm, 500 μm, 1000 μm, and 2000 μm, and different stripe length of 80 mm and 100 mm. The calculated thermopower of Pd/Cr TFTCs is 20.99 ± 0.1 μV/K, with a standard deviation of 8.27%. (b) Calibration results of 9 different Cr/Pt TFTCs. These TFTCs have the same size with the Pd/Cr TFTCs. The calculated thermopower of Pd/Cr TFTCs is 17.59 ± 0.3 μV/K, with a standard deviation of 1.12%.

Back to article page