Figure 5
From: Radiopaque Resorbable Inferior Vena Cava Filter Infused with Gold Nanoparticles

Infusion with gold nanoparticles had no significant effect on mechanical strength or integrity of PPDO sutures. The mechanical strength of the untreated and AuNP-infused PPDO sutures was tested by (A) the load-at-break method, (B) evaluating heat flux at various temperatures, and (C) melting temperature (T m ) of each sample by differential scanning calorimetry. No significant differences were noted on any of these parameters. The load-at-break values did not differ significantly between the two groups (Kruskal−Wallis one-way ANOVA on ranks, p < 0.05). The heat reflux curves superimposed on each other, and all measured T m values were around 112 °C.