Table 2 Reliability estimates of quantitative optical coherence tomography-angiography (OCT-A) metrics measurements using an automated analysis program in a subset of 30 eyes from 30 diabetic participants. Two OCT-A scans were acquired during first visit for these participants for intra-session repeatability assessment. These 30 patients were also invited for another visit for OCT-A scan within 2 weeks for inter-session reproducibility assessment.
 | Intra-section | Inter-section | ||
---|---|---|---|---|
ICC (95% CI) | CR (95% CI) | ICC (95% CI) | CR (95% CI) | |
FAZ Area (mm2) | 0.976 (0.950–0.988) | 0.040 (0.036–0.043) | 0.967 (0.932–0.984) | 0.050 (0.045–0.055) |
FAZ Circularity | 0.751 (0.540–0.873) | 0.125 (0.114–0.138) | 0.720 (0.490–0.856) | 0.129 (0.117–0.141) |
Total Vessel Density (%) | 0.840 (0.691–0.920) | 0.069 (0.063–0.076) | 0.777 (0.583–0.887) | 0.090 (0.081–0.098) |
Parafoveal Vessel Density (%) | 0.826 (0.667–0.913) | 0.078 (0.071–0.085) | 0.774 (0.577–0.885) | 0.098 (0.088–0.107) |
Fractal Dimension | 0.989 (0.978–0.995) | 0.010 (0.009–0.011) | 0.987 (0.972–0.994) | 0.010 (0.009–0.011) |
Vessel Density Index (mm) | 0.633 (0.358–0.807) | 0.001 (0.0008–0.001) | 0.624 (0.346–0.802) | 0.001 (0.0008–0.001) |