Table 4 Retinal Oxygen Saturation Values in Healthy Individuals across the Literature.

From: Development and Validation of Non simultaneous Retinal Image Acquisition–Based Retinal Oximeter

Authors

Sample size

Retinal arteriolar oxygen saturation (%)

Retinal venular oxygen saturation (%)

Age of the subjects (years)

Oximetry

Mean ± SD

Mean ± SD

Current study, 2016

45

94.4, IQR

55.0, IQR

21 (18–35)*

Nonsimultaneous retinal oximetry

Kim et al.10

11

90.3 ± 6.6

55.3 ± 7.1

Not mentioned

Nonsimultaneous retinal oximetry

Geirsdottir et al.11

120

92.2 ± 3.7

55.6 ± 6.3

47 (18–80)*

Oxymap simultaneous oximetry

Lasta et al.6

20

91.0 ± 9

48.0 ± 11

23 ± 4

Imedos simultaneous oximetry

Jani et al.8

61

90.4 ± 4.3

55.3 ± 7.1

44.1 ± 14.7

Oxymap simultaneous oximetry

Yip et al.13

118

93.6 ± 6.9

54.2 ± 6.9

54.6 ± 6.86

Oxymap simultaneous oximetry

Kashani et al.14

45

93.0 ± 7

65.0 ± 5

61 ± 9

HCTIS simultaneous oximetry

  1. HCTIS, hyperspectral computed tomographic imaging spectroscopy; IQR, interquartile range; SD, standard deviation. *Values in median (IQR).