Table 3 Results of McNemar’s test comparing the proposed new assay and dual-MAbs CLIA kit to the IHC (n = 41).
From: Development of an Affimer-antibody combined immunological diagnosis kit for glypican-3
Proposed assay | Total | dual-MAbs CLIA kit | Total | ||||
---|---|---|---|---|---|---|---|
+ | − | + | − | ||||
GPC3- IHC | + | 26 | 6 | 32 | 29 | 3 | 32 |
− | 0 | 9 | 9 | 9 | 0 | 9 | |
Total | 26 | 15 | 41 | 38 | 3 | 41 |