Table 5 The mean difference, paired t-test, 95% LoA and ICC for intraocular lens calculation differences between the two different operators based on the average method (from average of 3 consecutive readings from each operator) using OA-2000 swept-source optical coherence tomography in cataract patients.
Device Pairings | Mean Difference ± SD | P Value* | 95% LoA | ICC (95% CI) |
---|---|---|---|---|
SRK/T (Φ = 2.5) (D) | 0.00 ± 0.10 | 0.922 | −0.20 to 0.20 | 0.999 (0.998 to 0.999) |
Holladay 1(Φ = 2.5) (D) | 0.00 ± 0.12 | 0.962 | −0.23 to 0.23 | 0.998 (0.997 to 0.999) |
Hoffer Q(Φ = 2.5) (D) | 0.00 ± 0.13 | 0.993 | −0.25 to 0.25 | 0.998 (0.997 to 0.999) |
SRK/T (Φ = 3.0) (D) | −0.01 ± 0.09 | 0.461 | −0.18 to 0.16 | 0.999 (0.998 to 0.999) |
Holladay 1(Φ = 3.0) (D) | −0.01 ± 0.10 | 0.501 | −0.20 to 0.19 | 0.999 (0.998 to 0.999) |
Hoffer Q(Φ = 3.0) (D) | −0.01 ± 0.11 | 0.499 | −0.22 to 0.20 | 0.999 (0.998 to 0.999) |