Table 6 The mean difference, paired t-test, 95% LoA and ICC for intraocular lens calculation differences between the two different operators based on a single method (from the first reading from each operator) using OA-2000 swept-source optical coherence tomography in cataract patients.
Device Pairings | Mean Difference ± SD | P Value* | 95% LoA | ICC (95% CI) |
---|---|---|---|---|
SRK/T (Φ = 2.5) (D) | 0.00 ± 0.19 | 0.541 | −0.36 to 0.37 | 0.996 (0.993 to 0.997) |
Holladay 1(Φ = 2.5) (D) | 0.00 ± 0.21 | 0.092 | −0.41 to 0.41 | 0.995 (0.992 to 0.997) |
Hoffer Q(Φ = 2.5) (D) | 0.00 ± 0.23 | 0.565 | −0.44 to 0.45 | 0.994 (0.991 to 0.996) |
SRK/T (Φ = 3.0) (D) | 0.00 ± 0.15 | 0.853 | −0.30 to 0.31 | 0.997 (0.995 to 0.998) |
Holladay 1(Φ = 3.0) (D) | 0.01 ± 0.17 | 0.706 | −0.33 to 0.34 | 0.996 (0.994 to 0.998) |
Hoffer Q(Φ = 3.0) (D) | 0.01 ± 0.19 | 0.924 | −0.36 to 0.38 | 0.996 (0.994 to 0.998) |