Table 1 Intra- and inter-observer 95% and 99.73% limits of agreement (LOA) for K1, K2, and TCT with the Pentacam HR, Orbscan IIz, and Casia SS-1000 in healthy (n = 30) and keratoconic eyes (n = 30), using one or the mean of three repeat measurements.

From: Improving precision for detecting change in the shape of the cornea in patients with keratoconus

 

Device

Parameter

Intra-observer agreement

Inter-observer agreement

1 Measurement

1 Measurement

Mean of 3 measurements

SD

95% LOA

99.73% LOA

SD

95% LOA

99.73% LOA

SD

95% LOA

99.73% LOA

Healthy eyes

Pentacam HR

Kmax

0.15

0.35

0.51

0.21

0.41

0.63

0.11

0.20

0.31

K1 (D)

0.08

0.15

0.24

0.09

0.19

0.29

0.04

0.08

0.12

K2 (D)

0.09

0.17

0.27

0.13

0.23

0.37

0.07

0.11

0.18

TCT (μm)

5.07

11.37

16.64

5.24

10.96

16.61

2.70

5.18

7.99

Orbscan IIz

K1 (D)

0.21

0.99

1.58

0.59

0.99

1.61

0.30

0.61

0.82

K2 (D)

0.5

0.9

1.41

0.40

0.7

1.11

0.23

0.36

0.59

TCT (μm)

9.48

17.95

27.81

9.37

18.1

27.85

4.91

10.21

15.32

Casia SS-1000

K1 (D)

0.17

0.34

0.52

0.22

0.43

0.66

0.09

0.19

0.28

K2 (D)

0.19

0.37

0.57

0.22

0.44

0.67

0.10

0.29

0.35

TCT (μm)

1.33

2.14

3.53

1.78

2.8

4.65

1.18

1.85

3.07

Keratoconus

Pentacam HR

Kmax

0.79

1.49

2.32

0.81

1.37

2.21

0.46

0.78

1.25

K1 (D)

0.24

0.49

0.74

0.39

0.72

1.13

0.21

0.35

0.57

K2 (D)

0.40

0.8

1.21

0.22

1.27

1.93

0.22

0.38

0.61

TCT (μm)

5.9

12.19

18.33

6.98

13.71

20.97

3.39

6.4

9.93

Orbscan IIz

K1 (D)

0.57

1.23

1.82

1.76

3.63

5.46

1.22

2.67

3.93

K2 (D)

0.65

1.18

1.86

2.28

4.85

7.22

1.54

3.20

4.80

TCT (μm)

26.63

46.03

73.72

27.40

53.08

81.58

17.24

33.73

51.65

Casia SS-1000

K1 (D)

0.88

1.72

2.63

1.01

1.81

2.86

0.58

2.46

3.52

K2 (D)

1.55

2.89

4.50

1.69

3.48

5.23

0.76

1.39

2.19

TCT (μm)

4.98

10.13

15.3

9.9

21.70

31.99

7.08

14.69

22.06

  1. LOA = Limits of agreement, SD = Standard deviation of difference of 2 measurements.