Table 2 Summary of cut-off values for corneal shape change in keratoconic eyes based on one and of mean three repeated measurements, expressed as upper 99.73% LOAs with upper 95% CI.

From: Improving precision for detecting change in the shape of the cornea in patients with keratoconus

Device

Parameter

Cut-off for corneal shape change

1 Measurement

Mean of 3 measurements

Pentacam HR

Kmax

2.73

1.55

K1 (D)

1.63

0.70

K2 (D)

1.68

0.75

TCT (μm)

21.97

12.12

Orbscan IIz

K1 (D)

6.60

4.72

K2 (D)

9.46

5.79

TCT (μm)

99.30

62.80

Casia SS-1000

K1 (D)

3.77

2.04

K2 (D)

7.18

2.68

TCT (μm)

38.39

26.64