Table 2 Comparison of sweat cortisol estimated using chemiluminescence immunoassay and Fe2O3/CCY based cortisol immunosensor.
From: Carbon fiber based electrochemical sensor for sweat cortisol measurement
Samples | Chemiluminescence immunoassay (CLIA) method (ng/mL) | Fe2O3/CCY immunosensor | ||||
|---|---|---|---|---|---|---|
Measured (ng/mL)* | Added (ng/mL) | Found (ng/mL)* | RSD (%) | Recovery (%) | ||
1 | 23 | 23.71 | 50 | 75.24 | 3.403 | 102.07 |
2 | 24 | 27.81 | 50 | 77.51 | 3.874 | 99.62 |
3 | 28 | 41.62 | 50 | 95.48 | 4.064 | 104.21 |