Table 4 Mean differences and SD of differences between IOS and TremoFlo in the overall population.

From: Comparison of Forced and Impulse Oscillometry Measurements: A Clinical Population and Printed Airway Model Study

 

Overall Population

Mean Difference (IOS-TremoFlo)

Standard Deviation of the mean difference

p-value

95% CI

R5 (Kpa.s.L−1)

0.04

0.06

<0.0001

(0.032–0.050)

R20(19) (Kpa.s.L−1)

0.03

0.03

<0.0001

(0.021, 0.033)

R5-R20(19) (Kpa.s.L−1)

0.01

0.05

0.002

(0.005, 0.019)

AX (Kpa.L−1)

−0.81

1.20

<0.0001

(−0.770, 0.328)

X5 (Kpa.s.L−1)

0.02

0.06

0.122

(−0.001, 0.014)

X20(19) (Kpa.s.L−1)

0.06

0.03

<0.0001

(0.048, 0.060)

  1. Definition of abbreviations: R5: Resistance at 5 Hz; R20(19): Resistance at 20/19 Hz; R5-R20(19): Resistance at 5 minus 20(19) Hz; AX: Area of Reactance; X5: Reactance at 5 Hz; X20(19): Reactance at 20(19) Hz. p-values obtained from Wilcoxon signed rank test (paired test), 95% confidence intervals are for the median of the difference.