Table 5 Detection and quantitation limits and real air sample results obtained from different environments.

From: Multivariate Optimization of Tenax TA-Thermal Extraction for Determining Gaseous Phase Organophosphate Esters in Air Samples

 

Concentrations (ng m−3)

TEP

TPP

TNBP

TCEP

TCIPP

TDCIPP

TBOEP

TPHP

EHDPP

TEHP

aMDL (ng m−3)

0.1

0.2

0.5

0.3

0.4

0.4

0.5

0.5

0.5

0.3

bmLOD (ng m−3)

 

0.2

0.5

0.5

0.08

nd

nd

0.02

nd

0.4

aMQL (ng m−3)

0.4

0.5

1.6

0.9

1.3

1.2

1.7

1.5

1.8

1.1

bmLOQ (ng m−3)

 

0.5

1.6

1.8

0.3

nd

nd

0.07

nd

1.2

Office 1

<0.4

<0.5

6.5

0.05

0.3

<1.2

<1.7

<0.07

<1.8

<1.2

Office 2

<0.4

2.2

<1.6

1.3

<0.3

<1.2

<1.7

<0.07

<1.8

<1.2

Office 3

<0.4

1.6

2.0

2.3

<0.3

<1.2

<1.7

<0.07

<1.8

<1.2

Bedroom 1

<0.4

<0.5

6.3

2.7

0.3

<1.2

<1.7

<0.07

<1.8

1.6

Bedroom 2

<0.4

<0.5

<1.6

6.2

0.4

<1.2

<1.7

<0.07

<1.8

3.7

Car

<0.4

<0.5

2.8

1.6

<0.3

<1.2

<1.7

0.1

<1.8

<1.2

  1. aMDL and MQL are calculated by dividing the detection and quantification limits (LODs and LOQs) by the sampling volume.