Table 1 Sensing performance of tested taper profiles.

From: Detection of dengue using PAMAM dendrimer integrated tapered optical fiber sensor

Taper profile

Immersion time (mins)

Thickness (nm)

RMS

Sensitivity (nm/nM)

Average standard deviation (±)

P1

10

8.65

0.86

9.62

0.34

P2

20

12.36

1.05

14.60

0.75

P3

30

16.40

1.94

18.75

0.56

P4

40

21.10

1.95

19.53

0.51

P5

50

23.59

2.40

14.49

0.57

P6

60

48.42

4.69

13.99

0.51