Figure 7

From: A modular paper-and-plastic device for tuberculosis nucleic acid amplification testing in limited-resource settings

Figure 7

TB testing for clinical samples in FLIPP-NAAT. (A) Three clinical samples were tested per device with a negative and positive control in each module. Each sample was run in duplicates. Numbers on top of each zone represent the sample number and suffix ā€˜I’ and ā€˜II’ represent the duplicates. N – negative control and P – positive control. (B) Mean pixel intensity of green color from all the six blind clinical samples, negative and positive controls. The dashed line represents the threshold value of (µN + 3σN). Samples with intensity below the threshold were considered as TB negative while samples with green color intensity above the threshold were considered as TB positive.