Table 2 FGMS analytical accuracy as a function of various factors.

From: Clinical Performance of Flash Glucose Monitoring System in Patients with Liver Cirrhosis and Diabetes Mellitus

 

LC group

Control group

p value

MARD, %

95% CI

MARD, %

95% CI

Overall

12.68

12.14–13.22

10.55

10.06–11.02

<0.001*

First week

11.47

10.76–12.19

9.62

9.04–10.20

<0.001*

Second week

13.92

13.15–14.70

11.39

10.69–12.16

<0.001*

Below 80 mg/dL

11.52

6.35–14.68

8.62

6.45–10.74

0.950

80–120 mg/dL

12.51

11.78–13.24

9.37

8.61–10.20

<0.001*

Above 120 mg/dL

12.83

12.08–13.59

11.27

10.65–11.91

<0.001*

  1. LC: Liver Cirrhosis; MARD, %: mean absolute relative difference in percentage; 95% CI: Confidence Interval at 95% confidence level. *Statistically significant variables. All variables were analysed with Mann-Whitney test.