Table 5 Percent agreement between the reference device and the BB-613WP in identifying high and normal blood pressure. Hypertension was defined as a systolic value of > 140 mmHg or a diastolic value of > 90 mmHg. Kappa is given as a value (lower–upper limit).
Sensitivity (%) | Specificity (%) | Agreement (%) | Kappa | |
---|---|---|---|---|
Hypertension | 100 | 92.4 | 94.9 | 0.89 [0.79–0.98] |
No Hypertension | 100 | 96.2 | 96.5 | 0.74 [0.66–0.83] |