Table 4 (Ternary) CNN accuracy, sensitivity and specificity at the patient level of the ternary (COV+  vs COV− vs CTRL) classification model using leave-one patient out cross-validation, with the related COV+  sensitivity and specificity percentages. (Binary) CNNs accuracy, sensitivity and specificity at the patient level for the binary (COV+  vs CTRL, COV+  vs COV− and COV− vs CTRL) classification model.

From: COVID-19 salivary Raman fingerprint: innovative approach for the detection of current and past SARS-CoV-2 infections

 

Accuracy (%)

Sensitivity (%)

Specificity (%)

COV+  sensitivity (%)

COV+  specificity

Ternary

Patient-level

89

89

95

83.3

96

Patient-level (COV+  = 26)

92

91

96

85

96

Binary

COV+  vs CTRL

89

83

94

COV+  vs COV−

89.6

83

93

COV− vs CTRL

91.5

89

94