Figure 3 | Scientific Reports

Figure 3

From: Rapid, quantitative, and high-sensitivity detection of anti-phospholipase A2 receptor antibodies using a novel CdSe/ZnS-based fluorescence immunosorbent assay

Figure 3

Consistency analysis for QD-ICA and ELISA. (a) Comparison of Passing and Bablok regression analyses for the two assays (regression equation was Y = 0.91x − 0.38, the Spearman correlation coefficient r was 0.8). (b) Bland–Altman plot of log-transformed difference against the mean of QD-ICA and ELISA. Dashed lines represent 1.96 ± standard deviation (the confidence interval of repeated measures is 95%).