Table 1 Comparison of the fom obtained by cross-correlation and TDCR measurements with various levels of grey filters for \(^3\)H and \(^{14}\)C LSC sources.
From: Time-domain based evaluation of detection efficiency in liquid scintillation counting
\(^3\hbox {H}\): fom \(\varphi \), ph.e\(^-/\hbox {keV}\) | \(^{14}\hbox {C}\): fom \(\varphi \), \(\hbox {ph.e}^-/\hbox {keV}\) | ||||
---|---|---|---|---|---|
Filter | Cross-corr. | TDCR | Filter | Cross-corr. | TDCR |
None | – | 0.685(20) | None | – | 0.326(6) |
90% | 0.437(7) | 0.434(13) | 80% | 0.254(6) | 0.267(4) |
80% | 0.383(21) | 0.370(11) | 70% | 0.229(9) | 0.233(4) |
70% | 0.310(3) | 0.312(9) | 60% | 0.201(8) | 0.191(3) |