Table 1 Findings of cholangioscopy-guided balloon-based endobiliary radiofrequency ablation in an animal model.

From: Feasibility of balloon-based endobiliary radiofrequency ablation under cholangioscopy guidance in a swine model

Animal no.

Specimen no.

Sacrificed (day)

Bile duct site ablated

Balloon size

Target temperature

Ablation time

Impedance

Technical success

Confirmation of mucosal changes during the ablation on CS

Accuracy of the ablation area by CS

Complete internal circumference ablation

Length of ablation area (mm)

Depth of ablation area (mm)

Adverse events

1

1–1

1

Superior

6.0 mm × 20 mm

60 ℃

60 s

63 Ω

Yes

Yes

Yes

Yes

20.05

3.76

None

 

1–2

1

Inferior

6.0 mm × 20 mm

60 ℃

60 s

63 Ω

Yes

Yes

Yes

Yes

21.05

3.81

None

2

2–1

1

Superior

3.5 mm × 20 mm

(< 60 ℃)

92 Ω

Yes

No

Yes

No

Not obvious

0.21

None

 

2–2

1

Inferior

6.0 mm × 20 mm

(< 60 ℃)

96 Ω

Yes

No

Yes

No

Not obvious

0.84

None

3

3–1

1

Superior

6.0 mm × 20 mm

70 ℃

60 s

65 Ω

Yes

Yes

Yes

Yes

21.25

5.29

None

 

3–2

1

Inferior

6.0 mm × 20 mm

70 ℃

60 s

60 Ω

Yes

Yes

Yes

Yes

23.10

4.85

None

4

4–1

1

Superior

3.5 mm × 20 mm

60 ℃

60 s

68 Ω

Yes

Yes

Yes

Yes

20.07

3.10

None

 

4–2

1

Middle

6.0 mm × 20 mm

60 ℃

60 s

64 Ω

Yes

Yes

Yes

Yes

20.01

4.39

None

 

4–3

1

Inferior

6.0 mm × 20 mm

60 ℃

60 s

56 Ω

Yes

Yes

Yes

Yes

22.04

2.23

None

5

5–1

35

Superior

6.0 mm × 20 mm

60 ℃

60 s

80 Ω

Yes

Yes

-

Yes

19.05§

1.85§

None

6

6–1

35

Middle

6.0 mm × 20 mm

60 ℃

60 s

78 Ω

Yes

Yes

-

Yes

20.05§

3.02§

None

7

7–1

35

Superior

6.0 mm × 20 mm

60 ℃

60 s

67 Ω

Yes

Yes

-

Yes

21.00§

3.88§

None

  1. CS cholangioscopy.
  2. Time for which the target temperature was maintained.
  3. Did not reach the target temperature due to the thickness of the bile ducts and poor contact of the electrodes with the bile duct wall.
  4. §These are the areas of granulation tissue rather than the ablation (coagulative necrosis) area.