Table 1 Average Ct values obtained for three target regions from a series of 7 tenfold serial dilution of a clinical specimen previously detected SARS-CoV-2 positive (SC005G) using Meril covid-19 detection kit.

From: In-house reverse transcriptase polymerase chain reaction for detection of SARS-CoV-2 with increased sensitivity

Target 1 (RdRp)

Target 2 (E-gene)

Target 3 (S-gene)

Dilutions

Replicates detected

Average CT-values

SD

Replicates detected

Average CT-values

SD

Replicates detected

Average CT-values

SD

10–1

5/5(100%)

22.87

0.553968

5/5(100%)

23.67

0.915341

5/5(100%)

24.25

0.604276

10–2

5/5(100%)

25.77

0.663928

5/5(100%)

26.22

0.663099

5/5(100%)

27.62

0.934532

10–3

5/5(100%)

26.62

0.703314

5/5(100%)

27.92

0.557692

5/5(100%)

28.43

0.79117

10–4

5/5(100%)

31.42

1.406722

5/5(100%)

32.58

0.978673

5/5(100%)

33.45

0.41142

10–5

5/5(100%)

34.28

0.407799

5/5(100%)

33.77

0.721249

5/5(100%)

34.30

0.44238

10–6

2/5(40%)

37.42

1.286934

2/5(40%)

38.21

0.59397

3/5(60%)

39.53

0.600083

10–7

0/5(0%)

0/5(0%)

0/5(0%)