Table 3 Performance of ELISA based on Luminex diagnosis as gold standard.

From: Agreement between commercially available ELISA and in-house Luminex SARS-CoV-2 antibody immunoassays

Isotype

FN

FP

TN

TP

Total

SE (%)

SP (%)

PPV (%)

NPV (%)

IgG

22

1

121

27

171

55

99

96

85

IgM

1

112

24

34

171

97

18

23

96

IgA

3

102

22

44

171

94

18

30

88

All isotypes

3

73

20

75

171

96

22

51

87