Table 1 Comparison of AtoH sampling efficiency evaluated by monodisperse and polydisperse aerosol.

From: An experimental method for efficiently evaluating the size-resolved sampling efficiency of liquid-absorption aerosol samplers

Sampling

Flow Rate (LPM)

Aerodynamic Diameter (μm)

AGI-30

BioSampler

FPSL (%)

ATD (%)

P value

FPSL (%)

ATD (%)

P value

12.5

0.77 or (0.74 ~ 0.88)

(75.56 ± 5.28)

(78.69 ± 3.36)

0.44

(73.26 ± 1.42)

(73.66 ± 5.10)

0.91

1.1 or (1.01 ~ 1.21)

(81.25 ± 10.93)

(80.83 ± 0.49)

0.95

(82.91 ± 10.16)

(76.56 ± 3.41)

0.40

1.9 or (1.75 ~ 2.09)

(87.44 ± 7.11)

(84.73 ± 1.45)

0.58

(88.14 ± 7.25)

(80.87 ± 1.81)

0.21

9

0.77 or (0.74 ~ 0.88)

(61.55 ± 7.63)

(66.98 ± 1.08)

0.34

(60.28 ± 8.42)

(67.46 ± 5.33)

0.29

1.1 or (1.01 ~ 1.21)

(71.49 ± 6.78)

(70.72 ± 2.71)

0.87

(61.07 ± 8.42)

(71.41 ± 3.71)

0.16

1.9 or (1.75 ~ 2.09)

(77.13 ± 5.79)

(76.77 ± 4.52)

0.94

(74.88 ± 5.52)

(76.89 ± 0.90)

0.59

6

0.77 or (0.74 ~ 0.88)

(43.97 ± 9.14)

(51.31 ± 1.03)

0.30

(36.31 ± 9.15)

(48.40 ± 0.37)

0.15

1.1 or (1.01 ~ 1.21)

(56.06 ± 4.58)

(57.69 ± 2.16)

0.62

(49.87 ± 6.07)

(52.49 ± 0.86)

0.53

1.9 or (1.75 ~ 2.09)

(64.95 ± 4.41)

(65.75 ± 3.52)

0.82

(60.51 ± 4.75)

(58.03 ± 2.23)

0.48

  1. 0.77 or (0.74 ~ 0.88) in the table were aerodynamic diameter of FPSL as statement and ATD particle as measured by laser diffraction method. Mean ± SD was used to indicate the range of sampling efficiency. Welch Two Sample t-test was used to assess the difference between evaluation results of monodisperse and polydisperse aerosol, and p < 0.05 was considered statistically significant.