Table 2 Results of quantitative PET image assessment, including maximum standardized uptake value (SUVmax) of the lung nodules (n = 144), nodule signal-to-background ratio (SBR), nodule signal-to-noise ratio (SNR), contrast-to-background ratio (CBR), and contrast-to-noise ratio (CNR) in block sequential regularized expectation maximization (BSREM) compared to ordered subset expectation maximization (OSEM) reconstructions as reference.
BSREM | OSEM | p valuea | |
|---|---|---|---|
SUVmax | 0.0001 | ||
Mean | 5.4 | 3.6 | |
Median | 4.0 | 2.7 | |
Range | 1.2–20.7 | 0.7–17.4 | |
SBR | 0.0001 | ||
Mean | 3.3 | 2.2 | |
Median | 2.4 | 1.6 | |
Range | 0.7- 15.7 | 0.4–8.6 | |
SNR | 0.0001 | ||
Mean | 21.2 | 14.6 | |
Median | 15.9 | 10.5 | |
Range | 3.9–89.4 | 2.5–75.7 | |
CBR | 0.0001 | ||
Mean | 1.2 | 0.3 | |
Median | 0.54 | − 0.06 | |
Range | − 0.6 to 14.7 | − 0.8 to 4.5 | |
CNR | 0.0001 | ||
Mean | 7.1 | 2.1 | |
Median | 3.0 | − 0.4 | |
Range | − 4.9 to 67.1 | − 7.2 to 37.8 |