Table 3 Clinical event detection.

From: Clinical feasibility of an advanced neonatal epidermal multiparameter continuous monitoring technology in a large public maternity hospital in Nairobi, Kenya

 

High heart rate

Low heart rate

High respiratory rate

Low respiratory rate

Low oxygen saturation

High temperature

Pre

Post

Pre

Post

Pre

Post

Pre

Post

Pre

Post

Pre

Post

Event

True positive

109

109

4

4

1090

1111

3

3

535

1165

1

True negative

1426

1633

1661

1661

463

536

1126

1126

464

514

602

False positive

240

33

4

4

76

3

14

14

56

6

6

False negative

0

0

0

0

81

60

0

0

743

113

3

Statistical summary

Accuracy (%)

87

98

100

100

91

96

99

99

56

93

99

Sensitivity (%)

100

100

100

100

93

95

100

100

42

91

25

Specificity (%)

86

98

99

99

86

99

99

99

89

99

99

PPV (%)

31

77

50

50

94

100

18

18

91

100

14

NPV (%)

100

100

100

100

85

90

100

100

38

82

100

FN:FP

0

0

0

0

1:0.9

1:0.1

0

0

1:0.1

1:0.1

1:2

  1. Pre-adjudication (Pre); post-adjudication (Post); positive predictive value (PPV); negative predictive value (NPV); false negative event to false positive event ratio (FN:FP).
  2. Accuracy = (True positive + True negative)/(True positive + True negative + False negative + False positive); Sensitivity = True positive/(True positive + False negative); Specificity = True negative/(True negative + False positive); PPV = True positive/(True positive + False positive); NPV = True negative/(True negative + False negative).