Table 3 The measurement parameters of the three configurations.

From: The effect of particle size and water content on XRF measurements of phosphate slurry

 

Excitation energy (kV)

Current (µA)

Filter

Measurement time (S)

Elements measured

Config 1

50

100

Ag

100

As, Zn, U, Cu, Sr

Config 2

12

416

Al

120

K, Ca, Cr, Fe

Config 3

10

316

None

300

Al, Si, P, S, Cl