Table 1 Scalar translocation event evaluation after cochlear implantation (n = 32)⁑.

From: The effect of the surgical approach and cochlear implant electrode on the structural integrity of the cochlea in human temporal bones

Inter-observer agreement UCR*: reviewer 1 vs. reviewer 2

Inter-observer agreement MPR#: reviewer 1 vs. reviewer 2

Inter-method agreement: CB-CT vs. Histology

UCR† vs. Histology§

MPR† vs. Histology§

Observed agreement

к¶

Observed agreement

Šŗ

Observed agreement

Šŗ

Observed agreement

Šŗ

91%

0.85

69%

0.45

94%

0.89

74%

0.57

  1. ⁑scalar translocation event: minimal of one electrode-contact in scala vestibuli or scala media, including direct scala vestibuli insertions; *UCR uncoiled cochlear reconstruction CB-CT images; #MPR axial and sagittal multiplanar reconstructions; †if observations by reviewer 1 and 2 were different, consensus was achieved by final decision of reviewer 3; §based on observations of reviewer 3; ¶magnitude of kappa coefficient: < 0: poor, 0.00–0.20 = slight, 0.21–0.40 = fair, 0.41–0.60 = substantial, 0.81–1.00 = excellent agreement.