Table 1 Success rate of optimal rewiring and procedure time.

From: Importance of optimal rewiring guided by 3-dimensional optical frequency domain imaging during double-kissing culotte stenting demonstrated through a novel bench model

 

Angio-guidance N = 10

OFDI-guidance N = 10

p

After the first stent implantation

 Success of optimal rewiring at the first attempt, n(%)

5(50)

7(70)

0.325

 Success of optimal rewiring before KBI, n(%)

5(50)

10(100)

Rewiring point before KBI

  

 Proximal

4

0

 Distal (Optimal)

5

10

 Distal small

1

0

 Far-distal

0

0

 N of attempts, n

10

14

After the second stent implantation

 Presence of metal carina

8(80)

5(50)

0.0055

 Success of optimal rewiring at the first attempt, n(%)

1(10)

5(50)

 Success of optimal rewiring before KBI, n(%)

1(10)

8(80)

Rewiring point before KBI

  

 Proximal

7

2

 Distal (Optimal)

1

8

 Far-distal

0

 Abluminal distal

2

 No of attempts, n

12

15

 Success of optimal rewiring after both the first and the second stent implantation, n(%)

1(10)

8(80)

0.0055

 Procedure time, median (range), minute

25(22–64)

44(36–61)

0.0036

 Fluoroscopic time, median (range), minute

10.4(7.5–23.4)

15.1(8.9–23.3)

0.0376

  1. N,n; number, KBI: kissing balloon inflation.