Table 4 \(^3\)He detector optimisation input parameters and output dimensions and specific count rates.
From: A new ground level neutron monitor for space weather assessment
 | Iteration 1: | Iteration 2: | Iteration 3: | Iteration 4: |
---|---|---|---|---|
Spacing | Void thickness | Outer box thickness | Outer gap size | |
Spacing/cm | 0.6 | 0.6 | 0.6 | 0.6 |
Void thickness/cm | 1.25 | 0.3 | 0.3 | 0.3 |
Outer box thickness /cm | 7.5 | 7.5 | 3 | 3 |
Outer gap size/cm | 1 | 1 | 1 | 0.2 |
Moderator width/cm | 282.6 | 190.15 | 187.6 | 187.6 |
Moderator height/cm | 6.24 | 4.44 | 4.34 | 4.34 |
Specific counts/cnts s\(^{-1}\) kg\(^{-1}\) | 0.00744 | 0.00841 | 0.008760 | 0.00916 |