Table 1 Patient and scanner characteristics.

From: Assessment of image quality and impact of deep learning-based software in non-contrast head CT scans

 

Centre

p-Value (Site I vs. II, Site I vs. III, Site II vs. IIII)

Site I

Site II

Site III

Patient cohort

N

48

54

50

 

Male/Female (Percent)

20/28 (42%/58%)

28/26 (52%/48%)

27/23 (54%/46%)

0.41/0.31/0.98

Median Age (IQR)

63 (50; 77)

82 (68; 88)

73 (59; 83)

 < 0.001/0.01/0.01

Scan parameters

Reference mAs (A)/ Reference mAs (B)

300/300

273

Median mAs (A) (IQR)/ Median mAs (B) (IQR)

213 (202; 224)/ 226 (215; 239)

214 (204; 221)

230 (230; 230)

Median scan length in cm (IQR)

16.6 (15.8; 17.1)

16.6 (16.0; 17.5)

16.9 (15.8; 17.6)

0.59/0.30/0.60

Radiation dose

Median CTDIvol (IQR) (mGy)

33 (31; 34)

30 (29; 31)

48 (48; 52)

 < 0.001/ < 0.001/ < 0.001

Median DLP (IQR) (mGy∙cm)

528 (501; 583)

498 (469; 532)

847 (798; 847)

 < 0.001/ < 0.001/0.001

Median Effective Dose (mSv) (IQR)*

1.06 (1.00; 1.17)

1.00 (0.94; 1.06)

1.69 (1.60; 1.69)

 < 0.001/ < 0.001/0.001

  1. *Effective Dose was calculated using the E/DLP coefficient (0.002 mSv/mGy·cm) from Shrimpton et al.30.
  2. Significant values are in [bold].