Table 3 HR average values of macaques detected via ECG and radar.

From: Contactless vital signs monitoring in macaques using a mm-wave FMCW radar

Macaque ID

State

ECG

Radar

Abs. error

Rel. error (%)

16104

Awake

147.50

148.67

1.17

0.79

152024

Awake

145.83

144.50

1.33

0.91

152030

Awake

138.83

139.00

0.17

0.12

152032

Awake

150.50

150.17

0.33

0.22

15092

Awake

147.00

145.50

1.50

1.02

151010

Awake

160.00

159.67

0.33

0.21

151007

Awake

130.50

131.33

0.83

0.64

151009

Awake

128.83

130.17

1.34

1.04

19080

Anesthetized

86.34

86.50

0.16

0.19

19092

Anesthetized

101.67

100.83

0.84

0.83

20005

Anesthetized

85.50

84.00

1.50

1.75

20042

Anesthetized

78.00

77.83

0.17

0.22

20313

Anesthetized

87.00

86.67

0.33